View All Images
|
|
deltaCal Volumetric Air Flow Calibrator
The deltaCal is the first and only volumetric flow calibration / audit device which gives a direct indication of volumetric flow rate, barometric pressure, ambient temperature and filter temperature in one instrument.
|
|
|
|
|
|
|
|
deltaCal Volumetric Air Flow Calibrator Overview It was developed for use with U.S. EPA PM2.5 designated samplers and their PM10 derivatives. With an accessory adapter it is usable with any air sampler in the 2 to 20 lpm range. The Euro version has a range of 10 to 60 lpm.
The deltaCal is the only field audit / calibration device which meets the ame temperature criteria as the EPA designated samplers themselves. The temperature range is -30 to 40°C. The unit is NIST traceable and was designed and manufactured in an ISO 9001 facility.
The heart of the deltaCal is its venturi which is superior to an orifice because it offers no sharp edges which can be damaged and a 90% recovery of the measurement pressure drop - important when calibrating some forms of Speciation Samplers or other critical orifice controlled devices.
APPLICATIONS
- Calibration & audits of PM2.5 & PM10 samplers to U.S. EPA requirements
- May be used on any sampler in the 2 to 20 lpm range
- Usable in the field or laboratory
- Calibration of European PM10 samplers operating at 38.4 lpm
- Calibration of 50 lpm DICOT samplers
deltaCal Volumetric Air Flow Calibrator Specifications
| Flow Rate Range |
2 to 20 lpm (Euro 10 to 60 lpm) |
| Temperature Operational Range |
-30 to 40°C |
| Temperature Reading Range |
-40 to 50°C |
| Barometric Pressure Range |
400 to 800 mm of Hg (U.S. version); 534 to 1066 millibars (Euro version) |
deltaCal Volumetric Air Flow Calibrator Design deltaCal Volumetric Air Flow Calibrator Features
- One switch (off/on) operation - all data displayed on screen
- No moving parts
- 20-hour battery life (4-AA)
- Furnished with universal power supply (120/240 VAC)
- Supplied software with download & reporting capability
- New U.S. EPA PM2.5 & Chemical Speciation Sampler SOPs using the triCal calibrators
|
Low Flow, 0.1 to 30 lpm
Personnel at Federal, State and Local regulatory agencies will value the selection of air flow calibrators from New Star Environmental. Our air flow calibrators make it easy to ensure the task at hand is being completed properly and with accurate instruments. And every air flow calibrator is backed by 60 years of industry experiences, so you can feel confident in every purchase. For more information on our air flow calibrators, or to place an order, contact us today.
Back to Low Flow, 0.1 to 30 lpm
|
|
|
|