PM10 / PM2.5 Dichotomous Sampler
The PM10 / PM2.5 Dichotomous Sampler is a U.S. EPA Reference Method for PM10 Sampling under Designation Number RFPS-0789-073.
|
|
|
|
|
|
|
|
PM10 / PM2.5 Dichotomous Sampler Overview The "dichot" operates at a flow rate of 16.7 lpm (1 cmh) and is available with a built-in electronic digital programmable timer or mechanical 7-day skip timer. Separate control and sampling modules allow for flexibility in deployment.
The weather-resistant control module houses the diaphragm vacuum pump, precision flow meters, flow selector valves, vacuum gauges, flow controller and flow recorder. The PM10 inlet and virtual impactor of the sampling module ensure accurate particulate collection.
APPLICATIONS
- PM10 compliance monitoring
- PM2.5 fine particulate sampling
- Source discrimination
- Visibility monitoring
- No filter artifact formation
- XRF and complete elemental analysis
PM10 / PM2.5 Dichotomous Sampler Specifications
| Inlet Collection Efficiency |
Cut-point within 10±1 micron over wind speed of 2 to 24 kph |
| Elapsed Time Indicator |
xxxx.xx hours |
| Vacuum Gauges |
0 to 30in Hg vacuum |
| Environmental |
Meets EPA requirements; -20 to 40°C, 600 to 800 mm Hg |
| Exterior Materials |
Aluminum; anodized |
| Power Required |
100/115 VAC ±10%, 50/60Hz, 8 amp max; 230 VAC ±10%, 50Hz, 5 amp max |
| Overall Dimensions - Control Module |
16" H x 22" W x 11" D (41cm x 56cm x 28cm) |
| Sampling Module |
64" H x 30" dia. tripod base bolt circle (162cm x 76cm) |
| Net Weight |
Control Module - 60 lbs (27 kg)Sampling Module - 20 lbs (9 kg) |
| Mechanical Timer |
7-day timer, fully adjustable start & stop times in 15 minute intervals; ±30 min. accuracy |
| Digital / Timer Programmer |
All functions digital and quartz crystal controlled; ±2 min/wk accuracy |
| Constant Flow Controller |
±5% accuracy at 1 cmh over pressure drop range of 0 to 35 cm Hg across fine particle filter; has |
| Virtual Impactor Collection Efficiency |
Cut-point is 2.5 microns |
| Virtual Impactor Internal Losses |
Maximum value over range of 0 to 10 microns is < 2% |
| Flow Rate |
Total flow: 16.7 lpm (1 cmh)Fine particle flow: 15.0 lpm (0.9 cmh)Coarse particle flow: |
| Filter Media |
37 mm dia. Teflon |
| Filter Holder Cassette |
Polypropylene, 44.5 mm O.D. |
| Flow Meters |
±3% accuracy at above flows |
| Flow Event Recorder |
Pressure transducer type, 24-hour, 10 cm (4 |
| Vacuum Pump |
Diaphragm type, brushless motor, clean exhaust |
| Total Shipping Weight |
90 lbs (41 kg) |
PM10 / PM2.5 Dichotomous Sampler Design The design of the PM10 inlet allows only particles smaller than 10 microns to enter the virtual impactor, where they are accelerated at sufficient velocity to impact the particles larger than 2.5 microns into the virtual impactor receiver tube. These coarse particles impact through a void or "virtual" surface rather than onto a hard surface. The receiver tube has a flow rate of 1.67 lpm (0.1 cmh) for collection of the coarse particles; while the fine particles, less than 2.5 microns, follow a flow of 15 lpm (0.9 cmh) to the fine particle filter. PM10 / PM2.5 Dichotomous Sampler Features
- PM10 Inlet extensively tested and recognized as the standard for PM10 low volume sampling
- Virtual impactor is EPA single stage design with 2.5 micron cutpoint
- Precision flow rates, precision machining of the virtual impactor and a high-performance flow controller mean accurate PM10 & PM2.5 data
- Filter holders clearly marked "Coarse" & "Fine"
- Flow rates easily audited by a total flow audit adapter connected to an external low pressure reference flow device
- Automatic sample termination when proper flow levels cannot be maintained
|
Particle Instrumentation
Dependable Aerosol Particle Measurement
Regulatory personnel at the federal, state, and local level, as well as cleanroom managers and OSHA/NIOSH personnel rely on New Star Environmental for dependable particle measurement equipment. Our equipment meets all international standards for GIS and ISO standards for cleanrooms, and can be retrofitted to convert existing PM10 inlets to PM2.5 inlets. Plus, every piece of equipment is backed by 60 years of industry experience. So, when you need quality products and expert guidance, you need New Star.